Hardware Mode; Operating The Maxwell ® 16 Forensic Instrument; Setting Up A Run - Promega Maxwell 16 Forensic Technical Manual

Table of Contents

Advertisement

4.

Hardware Mode

The hardware mode chosen for each run depends on the purification protocol being used. Please refer to the Technical
Bulletin for the DNA IQ™ Reference Kit for Maxwell
information on the hardware mode setting required.
The existing hardware mode (SEV or LEV) setting is displayed in the upper right hand corner of the Home screen. The
instrument displays the hardware mode entered and does not sense the hardware installed. Always verify that the
hardware installed matches what is displayed on the Home screen. Make sure that the hardware and hardware mode
are appropriate for the purification kit being used. If the installed hardware is not correct for the kit being used, remove
the installed Magnetic Rod Assembly and replace it with the appropriate Magnetic Rod Assembly (see Section 7.A).
If the installed hardware does not match the hardware mode displayed, turn the machine off, and install the
appropriate hardware with the machine turned off. When the machine is powered up after the installation, the installed
hardware will match the hardware mode displayed on the Home screen.
5.
Operating the Maxwell
5.A. Setting Up a Run
Setting Up a Run in the SEV Mode
Verify that the Home screen indicates SEV and that SEV hardware is present. Select Run. If you configured the
instrument to record users, a list of the user names added will be shown. Select the appropriate user, and add the
matching PIN (if appropriate).
A Verification screen indicates which protocol was chosen.
1.
Figure 21. Verification screen.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com
16 or DNA IQ™ Casework Pro Kit for Maxwell
®
16 Forensic Instrument
®
16 for further
®
23
TM321 · Revised 4/15

Advertisement

Table of Contents
loading

Table of Contents