Glomax ® -Multi+ Detection System With Temperature Control - Promega E8051 Technical Manual

Glomax multi+ detection system
Table of Contents

Advertisement

8.
GloMax
®
-Multi+ Detection System with Temperature Control
A new feature found in the GloMax
shaker (Section 7) is the plate heating device. The GloMax
System Heater is capable of heating from ambient 20°C to 45° ± 0.75°C. Before a
run, allow ten minutes for the heater to warm up and calibrate to the desired
temperature. Access the Temperature Control option by selecting the "Tools"
button on the Instrument Control Screen. After two hours of inactivity, the
Temperature Control option will automatically shut off. If the temperature of
the instrument is different than the desired temperature when an assay run is
started, the GloMax
®
-Multi+ Detection System will display a warning window
with the option to continue with or cancel the run.
Important: The use of a plate cover film is required to prevent sample
!
evaporation and damage to the instrument.
8.A. Activating the Temperature Control Option
1. Touch the "Tools" button to go to the Tools screen (Figure 15).
2. Touch the "Activate Heater" button to select the heater option.
3. Enter the desired temperature using the keypad.
4. Wait at least 10 minutes for the heater to calibrate to the desired
temperature.
Figure 15. Accessing the Heater option via the "Tools" button.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
®
-Multi+ Detection System, along with the
®
-Multi+ Detection
Part# TM311
Page 21

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8031E8041E8061E9061E9031

Table of Contents