Changing The Tank Filter - DENTSPLY inLab MC X5 Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions inLab MC X5
B
C
D
64 78 759 D3608
D3608.201.01.07.02 10.2017
7.7.2

Changing the tank filter

NOTICE
Change the filter insert regularly
We recommend changing the filter insert monthly (REF 6387067).
If a message appears stating that the water pressure is too low, please
check whether the filter has become clogged or the tank filling level has
fallen below the "min" marking.
Use only filter cartridges approved by Sirona!
To change the filter, proceed as follows:
The unit is switched on.
No machining process is running.
NOTICE
Risk of damage to the hoses!
Pulling at the inlet and outlet hoses may damage them.
➢ Always draw off the hoses at the pipe piece.
1. Draw off the machine-sided tank inlet hose (A) with the large diameter
as well as the thin coolant supply hose (B) from the tank. Pay
attention to leaking residual liquid and catch it with a separate
collecting container, if necessary.
2. Open the tank lid clip (C) and pull the filter system out. Pay attention
A
to any remainder liquid.
3. Remove the filter cartridge using a separate collection container or
sink.
4. Install a new filter cartridge (REF 6387067) on the filter system. Make
sure of correct fit.
5. Change the coolant if necessary (see Replace the coolant when
using DENTATEC (for material classes PMMA/wax, composite and
grinding materials) [ → 82]).
6. Place the filter system in the tank.
7. Lock the tank with the tank lid clip.
8. Connect the tank inlet hose (A) with the large diameter as well as the
thin coolant supply hose (B) again to the tank system.
7 Maintenance and cleaning
7.7 Coolant system
81