Operating Instructions inLab MC X5
B
C
D
64 78 759 D3608
D3608.201.01.07.02 10.2017
1. Draw off the machine-sided coolant drainage hose (A) with the large
diameter as well as the thin coolant supply hose (B) from the tank.
A
2. Open the tank lid clip (C) and pull the filter system out.
NOTICE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not
permitted.
➢ Do not use any cleaning agents.
3. Fill the tank with water up to the mark (max, D) (approx. 10 liters).
NOTICE
Damage to surfaces!
When undiluted, DENTATEC grinding additive etches plastic surfaces
and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
4. Add approx. 250 ml DENTATEC to the tank. The mixture ratio is: 1l
water to 0.025l DENTATEC.
5. Place the filter system in the tank.
6. Lock the tank with the tank lid clip.
7. Reconnect the coolant drainage hose (A) with the large diameter as
well as the thin coolant supply hose (B) to the tank system.
5 Commissioning
5.2 Putting the unit into operation
43