7 Maintenance and cleaning
7.7 Coolant system
82
7.7.3
Replace the coolant when using DENTATEC (for material
classes PMMA/wax, composite and grinding materials)
NOTICE
Damage to the pump and the tool mandrel!
Too high a proportion of ceramic in the coolant will damage the pump
and the tool mandrel.
Change the coolant regularly!
We recommend changing the coolant after one month at the latest.
7.7.3.1

Preventing odors

All coolant additives contain a biodegradable preservative. Despite this,
however, odors may still develop under unfavorable conditions.
Observe the following:
● Change the coolant at least once a month.
● Drain the tank if you do not intend to operate the unit for more than
one week.
● Clean the tank if the odors recur.
● Add the coolant additive DENTATEC and fill the tank up to the brim
with water. Let it stand for at least 24 hours and then rinse it out
thoroughly with water once again.
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates
plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit.
➢ Do not spill DENTATEC.
NOTICE
Approved coolant additive
Use only DENTATEC as a coolant additive.
Dentsply Sirona
Operating Instructions inLab MC X5
64 78 759 D3608
D3608.201.01.07.02 10.2017