7 Maintenance and cleaning
7.10 Replacing the main fuse
102
7.10
Replacing the main fuse
WARNING
Electric shock
Disconnect the power plug at the unit end before replacing the fuses.
NOTICE
Fuse type
The main fuses F1 and F2 are of the T3.15A 500V type. They are
available under the REF 6361336.
➢ Use only fuses of the same type in the fuse holder!
✔
The power plug must be disconnected.
1. Use a screwdriver to carefully pry off the cover of the fuses on the
back side of the unit.
2. Pull out the fuse holder.
3. Replace the defective fuses.
4. Reinsert the fuse holder.
5. Close the cover.
Dentsply Sirona
Operating Instructions inLab MC X5
64 78 759 D3608