Error Messages And Measures - DENTSPLY inLab MC X5 Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions inLab MC X5
Error message
Error in the air circuit of the spindle sealing
air.
Error in the air circuit of the tool change
cylinder.
Error in the blast air circuit.
Error in the air circuit of the boost valve
(pressure increase).
The water pump reported an error.
The suction device reported an error.
The spindle reported an error.
Internal software error.
No system pressure was detected.
Unspecified communication error. Please
try again.
Time-out occurred during order processing
!
The suction device reported an error.
The suction device reported an error.
No valid calibration data available.
The door lock reported an error.
The cover of the instrument tray cannot be
closed.
The chuck cannot be opened.
The chuck cannot be closed.
No tool detected.
The tool could not be removed and is still
secure in the chuck.
The tool sensor reported an error.
64 78 759 D3608
D3608.201.01.07.02 10.2017
7.11

Error messages and measures

In the event of an error, the machine outputs error messages. Refer to the
following table for measures to be taken for the respective error.
Action
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Please check that there is enough water in the tank or that the inlet and
outlet are properly connected.
Please check that the suction device is connected and switched on or,
for wet processing, is switched off. Please also check that the plugs are
properly connected.
Please check the state of the tools. If the tools are heavily worn out,
replace them and re-start the process.
Please re-start the machine. If the error persists, contact customer
service.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Unspecified communication error. Please try again. If the error persists,
contact customer service.
Unspecified communication error. Please try again. If the error persists,
contact customer service.
Please check that the suction device is connected and switched on and
check that the plug is properly connected.
Please check that the suction device is switched off.
Please contact customer service.
Please check that the door is properly locked. If the error persists,
contact customer service.
Please check that the lock of the magazine is closed.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Please check the compressed air supply (7 bar). If the error persists,
contact customer service.
Please check that each tool is connected correctly.
Please try again to change the tool or change the tool manually using
the spindle maintenance in the machine configuration. Clean the chuck
using the accompanying maintenance kit.
Please clean the sensor and then manually actuate it a few times. If the
error persists, contact customer service.
7 Maintenance and cleaning
7.11 Error messages and measures
103