icon or by pressing the Foot Pedal.
6.
After completing the purge cycle, place a sterilized
30kHz Cavitron Ultrasonic Insert into the handpiece
and set the Power Level Control and Lavage Control to
your preferred operating position for ultrasonic
scaling.

BETWEEn PATIEnTS:

1. Remove the used Cavitron Ultrasonic Insert. Clean and
sterilize following the Infection Control Procedures that
were enclosed with your insert.
2.
Hold the handpiece over a sink or drain and activate
Purge function as described in Step 5 of the Start-Up
procedure.
3.
After the Purge cycle is complete, turn the System
to the OFF (0) position.
4.
Remove the Steri-Mate handpiece, clean and sterilize
following the procedures outlined in the Cavitron
Systems Infection Control Procedures booklet that was
enclosed with your system.
5. Before disinfecting surface of cabinets, lock the
touch screen by pressing the Lock icon. Disinfect the
surfaces of the cabinet, Power Cord, Handpiece Cable,
Tap-On Foot Pedal and Cable assembly (if applicable),
Water Supply lines as instructed in the Cavitron
Systems Infection Control Procedures booklet. DO
NOT SPRAY DISINFECTANT SOLUTION DIRECTLY ON
SYSTEM SURFACES*.
6. Inspect the handpiece cable for any breaks or tears.
7.
If using a closed water supply or DualSelect Dispensing
System, check for adequate fluid volumes for the next
patient.
8. When ready to use, press the unlock icon, place a
sterilized Steri-Mate Handpiece onto the handpiece
cable assembly and insert a sterilized ultrasonic insert
into the handpiece and adjust system controls as
preferred.
SHUT-DOWn PROCEDURES AT THE
EnD OF THE DAy:
Follow the "Between Patients" maintenance procedures,
Steps 1 through 6. In addition, it is recommended to close
the manual shut-off valve on the dental water supply system.
*NOTE: Broad spectrum hospital grade water-
based disinfection solutions are preferred, such as
Lysol IC. Some alcohol-based disinfectant solutions
may be harmful and may discolor
plastic materials.

10 2 weekly maintenance

It is strongly recommended that this system be
disinfected by chemically flushing the waterlines with
a 1:10 Sodium Hypochlorite solution (NaOCl) at the
end of each week. This can be accomplished by
connecting this device to the Cavitron DualSelect
Dispensing System or a number of other devices
available from your local distributors. Where this device
is connected to the Cavitron DualSelect Dispensing
System, please follow the DualSelect system's Directions
for Use manual. If connected to another device, please
follow those directions for use, keeping in mind that
a chemical flush should be performed at maximum water
flow for at least 30 seconds. The system should be left
undisturbed for 10 minutes but no more than 30 minutes
to allow the sodium hypochlorite solution to soak in the
lines. As a suggestion, it is recommended that a sign
be placed on the system stating that the SYSTEM IS
BEING DISINFECTED WITH A STRONG DISINFECTANT
AND SHOULD NOT BE USED. When ready, flush system
with clean water for at least 30 seconds or until sodium
hypochlorite odor disappears. ALL CHEMICALS MUST
BE FLUSHED FROM THE SYSTEM BEFORE IT IS READY
FOR PATIENT USE.

10 3 monthly maintenance

WATER LInE FILTER MAInTEnAnCE:
When the water line filter becomes discolored or the filter
icon illuminates, the filter should be replaced to prevent
reduced water flow to the Cavitron Touch system.
1. Verify that the system is turned OFF.
2. Disconnect the water supply hose from the dental
office water supply. If a quick disconnect connector is
attached to the end of the hose, relieve the water
pressure by pressing the tip of the connector in an
appropriate container and drain the water.
3. Grasp the fittings on either side of the filter disk and
twist counterclockwise. Remove the filter section
from either side of the water hose.
4. Install the replacement filter onto the water hose
fittings. The filter should be positioned to match up
with the correct hose fitting.
5. Hand tighten one hose fitting onto filter in a clockwise
direction. Tighten second hose onto filter in clockwise
direction. Reconnect the water supply line, operate unit
to bleed the air and test for leaks.
6. Turn the unit back ON, press and hold filter icon until it
disappears to reset filter maintenance icon.
13