the collet chuck and this should therefore be executed at the machine
because you can then grab the tool.
5.2.4
Removing the transport safety device and putting the
processing chamber into operation
The machine is equipped with a transport safety device that must be
removed during commissioning. To do this, proceed as follows:
1. Switch the unit on at the main switch.
2. Start the machine by pressing the On/Off button.
3. Open the door using the door release button.
4. Take out the transport safety device.
5. Insert the supplied sponge C into the drain groove D of the door.
6. Place the supplied screen insert B in the lower drain.
B
7. Connect the suction hose of the mandrel suction to the mandrel.
5 Commissioning
5.2 Putting the unit into operation