7.9 Procedure when processing Medentika-PreFace® abutment semi-finished products made of
2. Tighten the screw (A) again at the respective point.
3. Insert the fixing ring.
4. Separate the coolant tank from the machine for processing titanium
workpieces.
5. Connect the tank for cleaning (tank contents: water without coolant
additive).
6. Perform an automatic cleaning process (with a clamped Medentika-
®
PreFace
SW.
7. Wipe off the compartment. Concentrate in particular on any
accumulations of metal shavings at the front door seal.

Cleaning the drain sieve

7.9.3.2
➢ Empty the drain sieve after processing no more than 6
®
PreFace
abutments.
holder) via the service menu of the software inLab CAM
7 Maintenance and cleaning
99