● Do not use compressed air, ultrasound and a jet of steam to clean the
HF-mandrel. Dirt can penetrate into the storage area.
Proceed as follows to clean the area:
NOTICE
Do not interrupt the maintenance/changing collet chuck process!
If you have started the process once, do not cancel the operation. Do
not turn off the machine during the changing process and do not
disconnect the power supply to the machine.
1. Start the inLab MC X5.
2. Start the inLab CAM software.
3. In the system start menu click on
"Machine and Instrument tray Management"
4. Click
"inLab MC X5"
5. Click on the
"Service"
6. Click
.
"Spindle Maintenance "
7. Click
The machine moves into the position for spindle maintenance.
The dummy tool is stored in the slot provided.
The collet chuck is opened.
The sealing air is switched on.
8. Remove the suction guard.
9. Now rotate the collet chuck out of the mandrel with the intended tool.
10. Clean the inner cone of the shaft of the HF-mandrel with the felt cone
from the service kit. The inner cone must be free of chips and dirt.
Then, clean the tool cone with a clean soft cloth or a clean, soft brush.
11. Now clean the collet chuck and the collet chuck holder with a clean
soft brush as well as the cylinder brush with eyelet. Alternatively use
a new collet chuck for the next steps.
7 Maintenance and cleaning
7.4 Maintenance for the tool mandrel
"Configuration"
.
.
.
.