Hitachi TM3030 Plus Instruction Manual page 372

Tabletop microscope
Table of Contents

Advertisement

Figure 7.24.2-2 Calibration of Z Message Dialog 2
5. The adjustment dialog shown in Figure 7.24.2-3 is displayed. Input the calibration value
of height (Z) between two points, or the angle between two points. And then, click the
[Next] button.
NOTE : In case of inputting the angle, the value of height (Z) is calculated and calibrated
by the value of horizontal (X) and input angle.
6. The confirmation message shown in Figure 7.24.2-4 is displayed. When clicking the [OK]
button, Z calibration is executed. The left side of the [Z] button in the [Calibration] block
changes into an orange mark.
Z calibration ends with the above-mentioned procedure. When selecting the [Z] from the
[Calibration] menu or clicking the [Z] button on the [Calibration] block with Z calibration
executed, the content of Z calibration is cleared and returned to the default.
Figure 7.24.2-3 Adjustment of Z Dialog
Figure 7.24.2-4 Confirmation Dialog
7 - 78