5.5 Error Messages

5.5.1 Messages Displayed When the Application is Started

The error messages listed below are displayed right after the application is started.
<<Error Message List>>
Code
Error Message
1000
Failed to access the memory.
Could not start up the
application.
1001
Memory size is inadequate.
Could not start up the
application.
1004
No USB driver found.
Could not start up the
application.
1050
Could not start up the
application.
Please check "Display
Setting".
1080
Failed to access Initial File.
Could not start up the
application.
Countermeasure
Click the [OK] button to close the program.
When an access error occurs on the hard disk,
this message appears. Make sure PC is normal
and the application program has been installed
properly.
Click the [OK] button to close the program.
When free space in the hard disk is less than
100 MB, this message appears. Increase the
free space to 100 MB or more by deleting
unnecessary files.
Click the [OK] button to close the program.
Some of the files required to run this program
may be missing. Install the USB driver.
For installation of the USB driver, refer to 2.5.4
and 2.5.5 in the instruction manual.
Click the [OK] button to close the program.
This message appears when the display
resolution is not set at 1024 x 768 pixels or more
and the display color is set below 16-bit color
level. Change display setting to meet the
condition.
For display setting, refer to 2.5.6 in the
instruction manual.
Click the [OK] button to close the program.
Access to this program is not authorized. Log on
to the computer as an Administrator and start
the program.
5 - 29