Hitachi TM3030 Plus Instruction Manual page 269

Tabletop microscope
Table of Contents

Advertisement

Code
Error Message
1120
No User Setting File found.
Do you default to start up PC?
1121
No data found in User Setting
File.
Do you default to start up PC?
1122
The data in User Setting File
are wrong.
Do you default to start up PC?
1150
Time has run out.
Start up the main unit and
application again.
Countermeasure
User setting file "TM_User.ini" is not found. To
start this application in default setting, click the
[OK] button, and the [Cancel] button to terminate
this application.
User setting file "TM_User.ini" contains no data
or some of the data are missing. To start this
application in default setting, click the [OK]
button, and the [Cancel] button to terminate this
application.
User setting file "TM_User.ini" contains
abnormal data or some of the data are missing.
To start this application in default setting, click
the [OK] button, and the [Cancel] button to
terminate this application.
A time-out has occurred during a startup
processing. Make sure the USB cable is
connected properly and the main power supply
is turned on.
Restart the system inby the following procedure.
1. Turn off the PC power.
2. Turn off the main unit power.
3. Restart the PC.
4. Turn on the main unit power and start the
application.
If the same message appears repeatedly after
restart of the system, contact the nearest service
representative.