Commissioning
50
Parameter overview with brief description
Parameter
Select gas type
Select the gas type for the
measurement application.
10.4.2

Specifying the process pressure

Navigation path
"Setup" menu → Process pressure
Parameter overview with brief description
Parameter/
Process pressure
Process pressure value for
calculating pressure-dependent gas
properties
10.4.3

Determining the installation factor

Navigation path
"Setup" menu → Installation factor
Parameter overview with brief description
Parameter
Installation factor
The factor is multiplied by
the mass flow to correct
sub-optimum installations
Description
Selection/
User entry
Gas type choose list
• Air
• Argon Ar
• Carbon dioxide CO2
• Nitrogen N2
Description
0.5 to 41.6 bar a
(7.3 to 603 psi a)
Description
Selection/
User entry
0 to 9
Proline t-mass A 150 HART
Factory setting
Air
Selection/
Factory setting
User entry
Country-dependent:
1.0130 bar a
(14.692 psi a)
Factory setting
1
Endress+Hauser