Beckman Coulter Microtube Precautions
• Beckman Coulter Microtubes are designed for use on specific Synchron systems. Using the
appropriate Microtube is essential for proper system operation.
• The sample height in the Microtube is critical for correct sample aspiration on all Synchron
systems.
• The use of Array Microtubes (PN 448163 or PN 448162) on Synchron Systems or the use of
Synchron Microtubes (PN 756776) on Array systems may result in short sampling, incorrect
results, and/or sample probe damage.
• The use of non-Beckman Coulter, third party Microtubes, which have not been designed and
tested on Synchron Systems may result in system damage and/or short sampling.
Biohazard Precautions
All biohazard precautions should be observed when doing maintenance, service, or troubleshooting
on the system. Always wear appropriate personal protective equipment, and wash hands after
working on contaminated portions of the system.
Blood Barrier Collection Tube Precautions
When blood collection tubes that contain physical barriers are used, extra care should be exercised
to ensure that the barrier is tightly packed. Loose particles from the barrier could coat or plug the
sample probe, flow cell, chemistry modules, electrolyte injection cup (EIC), or cuvette wash station.
BUNm/UREAm Electrode Precautions
To prevent damage to the BUNm/UREAm electrode, do not insert the stir bar removal tool,
applicator stick, or any other object into the BUNm/UREAm cup unless the electrode has been
removed. Remove the electrode before you remove the stir bar.
CC Subsystems Priming Precautions
If any two of the CC Subsystem items are selected, all three are primed automatically. Make sure that
all three areas are ready to be primed.
CO
Membrane Replacement Precautions
2
Do not touch the membrane surface when installing the quad-ring.
Covers, Doors and Shields Precautions
To ensure optimum performance of the system, operate the system with reagent doors and all
shields and covers in place. To prevent possible motion errors, verify the proper positioning of any
removed and reinstalled cover or shield.
A13914AF
Safety Notice
Summary of Precautions