Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x05DC0200
CC Subsystem Error
4
0x05DC0201
CC Reaction Carousel
4
Subsystem Error
Event Log
CAU: Homing Error for CC
This is a generic homing error that
Subsystem
occurs whenever there is an issue with
the CC Subsystem. This will normally be
Message ID: 0x05DC0200
the second error or will be preceded by a
Subsystem: boot, homing
more definite error.
command
The parameter for the event log entry
will either show "InitPhotometer" if the
error occurred during an RA Results
subsystem SpinReactionWheel
command or will show "HomeRA" if this
error occurred while the RA Subsystem
was performing a startup home and
Prime.
CAU: Homing Error for CC
An error occurred while homing the
Reaction Carousel Subsystem
cuvette wash station or reaction
carousel. Check the Event Log (classes 2
Message ID: 0x05DC0201
and 4) for specific reaction carousel or
Subsystem: boot, homing
cuvette wash station related errors.
command
Description
• Press the STOP button and
• If the error continues, check Event Log for
specific subsystem errors.
• Contact Beckman Coulter Support Center for
additional information.
• Make sure the cuvette wash head is installed
properly. Refer to
wash head.
• Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
Home
• If the error continues after homing, contact
Beckman Coulter Support Center for additional
information.
Corrective Action
Home
the system.
Replace Cuvette Wiper
to view
the system.