Reagent Load/Calibration
Reagent Load
Figure 4.9 Reagent Carousel
1. Cartridge
2. Top Positions
3. Bottom Positions
5
To load reagent:
• When the screen prompts you to load the cartridge, open the reagent carousel door. Move
the cartridge in front of the bar code reader. A beep will indicate that the bar code is read.
• Check the screen to verify reagent data is added to the selected position. If the bar code
does not scan, manually type the information.
6
Close the reagent carousel door.
7
If multiple positions are selected to load or unload, repeat Step 4-6.
8
Before processing samples, calibrate and run QC if needed.

Reagent Parameters

Calibration parameters specific for a particular reagent lot are included with certain multipoint,
non-linear reagent cartridges. These parameters are referred to as Reagent Lot Specific Parameters.
The Reagent Lot Specific Parameters, contained in four bar codes, are included on a card in the
Reagent Kit.
These parameters:
• are loaded as part of a reagent load whenever the parameters are not found in the database.
4-10
1
2
3
A011871P.EPS
A13914AF