Install a new ISE drain pump tube:
• Connect the tube to the tube fitting and route it through the pinch clamp on one side of the
pump.
• Rotate the pump roller mechanism approiximately half a turn while feeding the tube under
the pump rollers.
• Route the tube through the second pinch clamp and attach it to the tube fitting on the other
side of the pump.
• Close the plastic cover.
6
Verify that the tube is properly positioned under the pump rollers and is not twisted.
7
Select
Close
8
Select
Exit F10
9
Select
1 Prime
10
Select:
All Hydropneumatic Subsystem
• Type
Start Prime
11
As the ISE primes, check for leaks or bubbles from the tube or fittings, ISE waste draining, and
the tube seated properly.
If you see a problem, select
starting again at Step 1 above.
12
Prime the ISE 20 times and watch for bubbles or leaks.
13
Select
Exit F10
A13914AF
to exit the maintenance procedure and return the system to Standby.
.
in the
5
Number of primes to repeat
.
Stop Prime
to return to the Utilities screen.
from the Prime dialog box.
field.
. Check the fitting connections or replace a damaged tube,
Maintenance
As-Needed/As-Required Maintenance
9
9-127