Safety Notice
Summary of Precautions
System Restore Precautions
System Parameter and Alignment data can be restored from the backup diskettes onto the system;
however, performing the Restore function deletes some or all files (depending on the areas
restored) from the hard drive.
Urine Sample Precautions
After analysis of ten consecutive urine electrolytes, run one replicate of electrolytes on Synchron
Calibrator Level 2 in the serum mode. This minimizes the potential for chloride drift due to matrix
effects of urine samples.