Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x02620012
CC Reagent Cartridge –
4
No Fluid Detected
Chemistry Name:
chemistry name
Reagent Position:
reagent position
Component:
component
0x02620017
CC Reaction Wheel
2
Subsystem Motion Error
Event Log
CAU: CC Reagent Cartridge No
Two consecutive reagent level sense
Fluid Detected
errors for same cartridge – cartridge is
unusable.
Message ID: 0x02620012
Subsystem: RASched
Chemistry Name:
Reagent Position:
Component:
CAU: CC Reaction Wheel
This error will occur with any type of
Subsystem Motion Error
motion error within the Reaction Wheel.
Message ID: 0x02620017
Subsystem: RASched
Action Name: action name
Description
• Remove bubbles from cartridge neck or floating
on reagent surface.
• Reload reagent cartridge.
Refer to
(CC).
• Raise canopy and check for obvious obstruction
of the Reaction Wheel.
• Make sure sample and Reagent Mixer paddles
are properly seated.
• Make sure Wash Tower is properly seated and
aligned to cuvettes.
• Press the STOP button and
• Contact Beckman Coulter Support Center for
additional information.
Corrective Action
Load/Unload Cartridge Chemistries
Home
the system.