Beckman Coulter UniCel DxC 600 Instructions For Use Manual page 313

Unicel dxc synchron clinical systems
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

3
Center the membrane retainer (1) over the membrane clamp (3). Press the membrane retainer
into the clamp, thus securing the membrane (2).
Figure 9.57
4
Place the retainer assembly upside down on the work surface so the clamp is on top and retainer
on the bottom. Grasp the electrode carefully and press it firmly against the membrane and into
the membrane clamp.
Figure 9.58
A13914AF
Maintenance
As-Needed/As-Required Maintenance
9
9-77