Beckman Coulter UniCel DxC 600 Instructions For Use Manual page 431

Unicel dxc synchron clinical systems
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

Table 12.9 Result Errors and Codes (Continued)
Result Displayed on Report
"Result Printed"
"Result Printed"
"Result Printed"
"Result Printed"
"Result Printed"
<X (X=lower inst range)
<X (X=lower reportable range)
>Y (Y=upper inst range)
>Y (Y=upper reportable range)
CAL FAILED
CAL TIMED OUT
CHEM ABORTED
CHEM BYPASSED
CHEM NEEDS CAL
CHEM NOT RUN
DAC ERROR
DIL1 LEV SENSE
DIVISION ERROR
HI REAGENT
INCOMPLETE
INSUFF DIL1
INSUFF REAG
INVALID CHEM
LEV SENSE ERR
LO REAGENT
MOTION ERR
NO DIL1 ON SYS
NO REAG ON SYS
NO SAMP DETECT
NO SAMP ON SYS
NOT ACCEPTED
PENDING TEST
PROBE OBSTRUCT
A13914AF
Printed in
Host
Remarks Area
Code
CRITICAL HIGH
CRITICAL LOW
HIGH
LOW
TEMP ERR
TM
LT
LT
GT
GT
ZD
AB
Troubleshooting Calibration and Result Errors
Error Codes and Definitions
Inst.
Description
Code
Exceeds Critical range high
Exceeds Critical range low
Exceeds reference range high
Exceeds reference range low
T
Temperature error
Less than X - OIR LO
Less than X - ORR LO
Greater than Y - OIR HI
Greater than Y - ORR HI
Calibration failed
Calibration timed out
Too many retries
Channel bypassed
Chemistry needs calibration
Chemistry not run
ISE DAC Failure
Level sense error in diluent cartridge
Division Error Special calculation divisor is
zero
MC reagent level high
Calculation incomplete
Insufficient diluent reagent for scheduled
tests
Insufficient reagent strength
Invalid chemistry
Low reagent fluid
MC reagent level low
Motion error
No diluent reagent on the reagent carousel
Reagent not on board
Low sample fluid
Sample not on board
Wrong state for sample program message
Not ready
Probe obstruction detected
12
12-15