Beckman Coulter UniCel DxC 600 Instructions For Use Manual page 281

Unicel dxc synchron clinical systems
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

3
The syringe rods (1) are now positioned for placement.
Figure 9.29
4
Unscrew the barrel of the reagent syringe (left back wall) or barrels of two sample syringes
(back wall above modular cups) to release each syringe from the Luer-lock fitting. Refer to
Figure
9.28.
5
Unscrew the round plunger end (1).
Figure 9.30
A13914AF
1
A015944P.EPS
1
A011874P.EPS
9
Maintenance
Three-Month Maintenance
9-45