Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x13A00059
Scanned barcodes in
10
rack #<rack ID>
contain invalid
characters. Rack will be
unloaded from <rack
location>
0x13A0007E
Module: xxxxxx is
10
disabled due to a
system error. All
module functioning has
now ceased. When the
problem is fixed, enable
the module through
Instrument Commands.
0x13A00092
Sample Programs are
10
too old on rack: rack
number
Position Sample ID
1 sample ID1
2 sample ID2
3 sample ID3
4 sample ID4
Event Log
WLM: Barcodes contain invalid
Sample bar code contains illegal
chars.
characters.
Rack: <rack ID>
[1:<sample ID>]
[2:<sample ID>]
[3:<sample ID>]
[4:<sample ID>]
WLM: Received request to
Specified module has been disabled
disable module: Module (File
automatically. This will normally have
Name:Line Number)
severity of IM.
WLM: Sample programs too
Sample programming exceeds age
old
specified in setup. Usually indicates host
has reused sample IDs and the old
Rack: rack number
information was not cleared from the
[p: sample ID1]
instrument.
[p: sample ID2]
[p: sample ID3]
[p: sample ID4]
Description
• Check the sample bar code for illegal
characters. Refer to
• Make sure the complete bar code is visible and
there are no smudges or wrinkles. Refer to
Code
• Enable the module under Instrument
Commands,
problem continues, it could be a valve or
vacuum problem.
• Clear the existing sample program and
reprogram or requery host. Refer to
Samples.
Corrective Action
Valid Characters for
Entries.
Labeling.
Enable/Disable
Modules. If the
Clear
Bar