Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x04B20026
Sample tube did not
4
vent correctly. Remove
cap and reload tube.
Sample ID: sample id
Rack: sample rack id
Position: sample
position, MC Sample
Probe
0x04B20027
Sample tube did not
4
vent correctly. Remove
cap and reload tube.
Sample ID: sample id
Rack: sample rack id
Position: sample
position, MC Sample
Probe
0x04B3000F
ISE DAC Setting Failed
4
Chem Name: chem id
0x04B30010
MC Reagent Level High
4
in Reaction Cup
Chem Name: chem id
Event Log
CAU: MC Sample Tube Excess
Probe detected excessive vacuum in
Vacuum
thick stopper closed tube on second
attempt. Sample is aborted.
Message ID: 0x04B20026
Sample ID: sample id
Rack ID: sample rack id
Pos: sample position
CAU: MC Sample Tube Excess
Probe detected excessive pressure in
Pressure
thick stopper closed tube on second
attempt. Sample is aborted.
Message ID: 0x04B20027
Sample ID: sample id
Rack ID: sample rack id
Pos: position
CAU: ISE DAC Setting Failed
ISE DAC Setting Failed for the specified
chemistry.
Message ID: 0x04B3000F
Chem Name: chem id
CAU: Reagent too full
MC sample probe detected reagent in
the cup module above the expected
Message ID: 0x04B30010
level.
Cup: chem id
Description
• Uncap tube and rerun.
• If occurs consistently for many samples, check
cap piercer for broken blade. Refer to
CTS Blade/Wick (1-Blade Thick CTS Option)
Replace CTS Blade (1-Blade Narrow CTS Option)
to inspect the blade.
• Perform MC ODC Alignment.
• Uncap tube and rerun.
• If occurs consistently for many samples, check
cap piercer for broken blade. Refer to
CTS Blade/Wick (1-Blade Thick CTS Option)
Replace CTS Blade (1-Blade Narrow CTS Option)
to inspect the blade.
• Perform MC ODC Alignment.
• Make sure the ISE module is in the down
position, the cover is in place, the reagent door
is closed.
• Check for reagent volume in bottles or flow cell
leaks.
• Check electrode connections.
• Prime cup with reagent and make sure it drains.
Refer to
• Failure to drain is either caused by low vacuum
or a valve failure.
• Contact Beckman Coulter Support Center for
additional information.
Corrective Action
Replace
or
Replace
or
Priming MC
Subsystems.