Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x01FA002B
Hydro Smart Module
4
Communication
Problem. Stop/Home
may resolve this issue.
0x01FA0034
Hydro Illegal Switch
4
Condition
0x025C0001
Instrument Computer
4
Error
0x025C0009
Instrument Computer
4
Error
0x0262000F
CC Sample Probe
4
Disabled Due To Level
Sense Errors
Event Log
CAU: Hydro SM not
System stopped. Communication issue
communicating with MSMC.
with hydro smart module.
Check fiber optics
communication loop
Message ID: 0x01FA002B
CAU: Hydro Illegal Switch
Float Switch sending full and empty
Condition
messages at the same time.
Message ID: 0x01FA0034
Container: container
CAU: Debug port abort
Likely cause is the Instrument computer
has attempted to access a restricted
Message ID: 0x025C0001
area or file.
Subsystem: CAU
Problem Area: problem area
CAU: Bus Error
Likely cause is the Instrument computer
has received an invalid or unexpected
Message ID: 0x025C0009
response. This is not caused by user
Subsystem: CAU
input.
Task Name: task name
Vector ID: vector id
CAU: CC Sample Probe
Five consecutive sample level sense
Disabled Due To Level Sense
errors – probe, bead, bad alignment or
Errors
level sense board problem.
Message ID: 0x0262000F
System Response = Pause RA.
Subsystem: RASched
Description
• Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
Home
• If the error continues, contact Beckman Coulter
Support Center for additional information.
• Unscrew the canister. Wipe the float switch and
make sure the float is not stuck.
• Replace the Float Switch in the specified
container.
• At the Red Message screen, select Shutdown.
• If the problem continues contact Beckman
Coulter Support Center for additional
information.
• At the Red Message screen, select Shutdown.
• If the problem continues contact Beckman
Coulter Support Center for additional
information.
• Make sure Samples are in the correct rack.
• Perform CC Sample Probe to Cup and CC Sample
Probe Height alignments. Refer to step 3 of the
Replace the Old Probe
• Perform CC Sample Probe Level Sense
Alignment (Calibration).
Corrective Action
the system.
procedure.