Maintenance
Monthly Maintenance
5
Press the latch, which is located on the connector (1), to disengage the electrode pin lead from
the connector panel and remove the electrode connector from the socket.
Figure 9.13
6
Unscrew the BUNm/UREAm electrode retainer nut (1) and withdraw the electrode-retainer
assembly from the chemistry reaction cup.
Figure 9.14
7
Separate the electrode (3), retainer (2) and retainer nut (1). Refer to
It is not required to clean the stir bar, however, the stir bar may be cleaned if desired while the
BUNm electrode is removed from the cup. Refer to Four Month Maintenance,
Reagent
9-26
Lines, section in this chapter.
Figure
9.15.
Clean the MC
A13914AF