Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x05DC0096
Smart Module
4
Download Failed. Smart
Module ID: sm id
0x05DC01F8
CC Reagent Mixer
4
Motion Error
Event Log
CAU: Boot Task - SM Download
Smart module communication error
Failed
Message ID: 0x05DC0096
Subsystem: boot
SM ID: sm id
CAU: Homing Error for CC
This error occurs when there is a motion
Reagent Mixer
error with the Reagent Mixer, either
Rotary or Vertical. The parameter in the
Message ID: 0x05DC01F8
event log entry will show one of the
Subsystem: boot, homing
following:
command
• "AttachRgtMixer" – error occurred
• "RecoverRgtMixer",
• "HomeRgtMixerVer" – error occurred
• "EnergizeRgtMixerCrn",
• "HomeRgtMixerCrane" – error
Description
attaching to smart module 0xC1.
"EnergizeRgtMixerVer",
"RaiseRgtMixer" – error occurred
during a vertical movement toward
the home sensor flag.
during a vertical movement away
from the home sensor flag.
"MoveRgtMixerToSensor" – error
occurred during a horizontal
movement toward the hardware
sensor.
occurred during a horizontal
movement.
Corrective Action
• Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
Home
the system.
• If the error continues after homing,
the instrument and reboot.
• If the problem continues after the instrument
has been shutdown, contact Beckman Coulter
Support Center for additional information.
• Raise canopy and check for obvious obstruction
of the CC reagent mixer.
• Press the STOP button and
Home
While system is homing, observe motion of the
CC reagent mixer. The mixer should complete
its operations in a smooth manner.
• If the error continues, check the event log to
determine the specific subsystem error.
• Perform the appropriate vertical or horizontal
CC reagent mixer alignment.
• Contact Beckman Coulter Support Center for
additional information.
Shutdown
the system.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Unicel dxc 600iUnicel dxc 800

Table of Contents