Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x01FA001B
17 psi air supply
4
pressure is low
0x01FA001C
17 psi air supply
4
pressure is high
Event Log
CAU: 17 psi Air Pressure is Low
This error will occur whenever there is a
problem within the high-pressure
Message ID: 0x01FA001B
system. If the high pressure reads lower
than 14 psi for more than 15 seconds it
will be flagged as low.
CAU: 17 psi air pressure is high
This error will occur whenever there is a
problem within the high-pressure
Message ID: 0x01FA001C
system. If the High pressure reads
higher than 19 psi for more than 15
seconds it will be flagged as "High."
Description
Home
• If the instrument returns to Standby, adjust
pressure levels if necessary as follows:
— Make sure the 25 psi (Main Pressure) gauge
on the hydro reads within 24 to28 psi.
— Adjust the 25 psi (Main Pressure) regulator
as necessary.
— When the 25 psi pressure is within range,
make sure the High Pressure reading on the
Hydro Status screen is 17 ± 1.
— Adjust the 17 psi (High Pressure) regulator
as necessary.
— When the 17 psi pressure is within range,
make sure the Low Pressure reading on the
Hydro Status screen is within range.
— Adjust the 10 psi (Low Pressure) regulator as
necessary.
• If the instrument will not return to Standby,
contact Beckman Coulter Support Center for
additional information.
• Check pressure levels and
• Adjust pressure levels if necessary as follows:
— Make sure the 25 psi (Main Pressure) gauge
on the hydro reads within 24 to28 psi.
— Adjust the 25 psi (Main Pressure) regulator
as necessary.
— When the 25 psi pressure is within range,
make sure the High Pressure reading on the
Hydro Status screen is 17 ± 1.
— Adjust the high pressure system to 17 psi.
• Contact Beckman Coulter Support Center for
additional information.