Maintenance
As-Needed/As-Required Maintenance
8
Reinstall the cleaned drain tube, or install a new drain tube (PN 472698). Insert the drain tube
down through the O-ring until it sits flush on the bottom of the ISE drain.
NOTE
If the fit is too snug to insert the drain tube, it may be necessary to lubricate the O-ring on the
inside of the lower drain. Without removing the O-ring, apply a thin coat of Silicone compound to the
O-ring area inside the lower drain, prior to inserting the drain tube.
9
Clean the ISE drain top using a lintless tissue soaked with deionized water. Thoroughly wipe dry
the ISE drain.
IMPORTANT
top.
10
Reinstall the ISE drain top onto the drain. Align the two posts extending from the bottom of the
drain top, with the holes in the acrylic drain mount, and gently lower the drain top until it sits
flush (with no gaps) on the mount.
Tighten the screw at drain top.
11
Select
OK
12
After priming is complete, select
screen.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
13
Lower the ISE module by releasing the locking pin. Be careful NOT to pinch tubing as the module
is lowered.
14
Reinstall the ISE module cover.
15
Calibrate all the ISE chemistries.
9-90
Do not pinch, damage, or disconnect the adjacent connector wires while reinstalling the drain
to exit the Maintenance pop-up confirmation window when procedure is complete.
Close
in the pop-up
window to exit the
Results