Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries (Continued)
Type
Nonblanked Rate
Chemistries
A13914AF
FOR HIGH CALIBRATOR LEVEL:
Reaction Rate = Rate
rep1
Reaction Rate = Rate
rep2
(Rate
+ Rate
) × 0.5 = Rate
rep1
rep2
FOR LOW CALIBRATOR LEVEL:
Reaction Rate = Rate
rep1
Reaction Rate = Rate
rep2
(Rate
+ Rate
) × 0.5 = Rate
rep1
rep2
Cal Set Point (hi) – Cal Set Point (lo)
Cal Factor (Slope) =
Offset = Cal Set Point (hi) - [Cal Factor × Rate
Sample values are calculated by the following equation:
(Reaction Rate × Cal Factor) + offset = sample value
IMPORTANT
For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
Cartridge Chemistry: Calibration Theory
Formula
(hi)
avg
(lo)
avg
Rate
(hi) – Rate
(lo)
avg
avg
(hi)]
avg
System Description
E014418L.EPS
1-47