Maintenance
Monthly Maintenance
4
Refer to
• Fully loosen the one captive blade screw (2).
• Fully loosen the two captive blade holder screws (1).
• Remove the blade holder.
Figure 9.20
5
Hold the blade at the top of its shaft. Do NOT touch the points of the blade.
6
Lift up the blade to remove it.
7
Discard the old blade into a biohazard sharps container.
NOTE
The points of the blade are very sharp.
8
Remove a new blade from its container.
9
Hold the blade holder with the blade hole on the top. Note that the blade hole is keyed and the
blade has a slot.
Install the blade into the blade holder as follows:
• Rotate the blade, if necessary, until it falls into place.
• Finger tighten the captive blade screw.
• Remove the protection from the points of the blade.
• Do NOT touch the points of the blade.
9-34
Figure
9.20.
1
2
A014850P.EPS
A13914AF