Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x13A00024
QC Database is full. This
10
result was not saved.
0x13A0003F
Incomplete set of
10
calibrator levels in rack
rack ID for the following
calibrators: calibrator
name list
0x13A00056
No sample
10
programming or
pending test found for
samples in rack #<rack
ID>. Rack will be
unloaded from <rack
location>
0x13A00057
Required levels for
10
calibrator not loaded.
Rack will be unloaded
from the sample
carousel.
Rack: rack ID
Calibrator: calibrator
name
Event Log
WLM: QC DB is full. The result
QC database is full and there is no space
for
for the result to be saved. Results may
only overwrite data for existing,
control: control name /
previously defined control and chemistry
Chem: chem name has not
information.
been saved
WLM: Missing calibrator levels
Not all calibrator levels were loaded or
were loaded out of order.
Rack: rack ID
Calibrators: calibrator list (File
Name:Line Number)
WLM: No sample programming
A rack was loaded with nothing to run.
or pending tests found for
The rack is off-loaded and any previous
sample on rack.
results are printed.
Rack ID: <rack ID> (<File
Name>:<Line Number>)
WLM: Subsequent calibrator
Not all calibrator levels were loaded or
levels not loaded for multi-rack
were loaded out of order.
calibrator on rack rack ID (File
Name:Line Number)
Description
Archive QC
• Reload calibrator racks.
• Make sure bar codes are readable. Refer to
Code
• If there are 4 calibrators or less, they must all be
in the same rack.
• If more than 4 levels are required, levels 1–4
must be in the first rack loaded and the next
rack loaded must contain the remaining levels.
• Check the sample status. It may have been
completed and reloaded. Verify the sample
programming.
• Could indicate a problem receiving sample
programming from the LIS.
• Reload calibrator racks.
• Make sure bar codes are readable. Refer to
Code
• If there are 4 calibrators or less, they must all be
in the same rack.
• If more than 4 levels are required, levels 1–4
must be in the first rack loaded and the next
rack loaded must contain the remaining levels.
Corrective Action
data.
Labeling.
Labeling.
Bar
Bar