Advertisement

5
Refer to
Remove the blade and wick assemblies as follows:
• Hold the blade (1) at the top. Do not touch the points (2) of the blade.
• Lift up the blade and remove it with the wick assembly at the same time:
— The wick assembly (3) should be attached to the blade.
— Do not touch the wick because it contains CTS Auto-Gloss lubricant which is very
• Discard the old assemblies into a biohazard sharps container.
Figure 9.24
6
Refer to
NOTE
A new blade is shipped with a wick attached. The points of the blade are very sharp. The blade
shaft is not sharp.
• Remove a new blade and wick from its container.
• Hold it at the top of the blade (1), immediately under the blade clamp as shown in
Figure
A13914AF
Figure
9.24.
slippery and difficult to clean.
1
3
2
Figure
9.24.
9.24.
Two-Month Maintenance
A011872P.EPS
9
Maintenance
9-39