Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x02620025
CC Sample Syringe
2
Motion Error
0x02620033
10 Cuvettes Have failed
4
Water Blank
0x02620034
CC Reagent probe
4
disabled due to level
sense errors
Event Log
CAU: CC Sample Syringe
This error occurs when there is an error
Motion Error
within the Sample Syringe Subsystem.
This includes both the Shear valve Motor
Message ID: 0x02620025
and the Lead Screw Motor.
Subsystem: RASched
Action Name: action name
CAU: 10 Cuvettes Have Failed
This error can be flagged for the
Water Blank
following:
Message ID: 0x02620033
• A cuvette's wavelength absorbance
Subsystem: RaSched
• A cuvette's wavelength absorbance
• PSM Error.
• LPIA Failure.
CAU: CC Reagent Probe
Multiple reagent cartridges have level
Disabled Due To level Sense
sense errors.
Errors
Message ID: 0x02620034
Subsystem: RASched
Probe: probe
Description
is over 350 mA (Absorbance) for all
wavelengths between 340 to 700
nanometers.
is over 200 mA (Absorbance) for the
940 nanometer wavelength (PRO
System only).
Corrective Action
• Make sure syringe and plunger are properly
seated at top and base of syringe.
• Check for obstruction of syringe movement.
• Perform CC Syringe Home Alignment.
• If the error continues, contact Beckman Coulter
Support Center for additional information.
• Remove wash station head and inspect probes
for blockage and wiper for wear. Refer to
Replace Cuvette Wiper
for removal of Wash
Station.
• Check wash station wiper alignment.
• Check wash station alignment.
• Remove reaction carousel and check cuvette
cleanliness.
Non-resolution of problem may indicate photometer
issues. Contact Beckman Coulter Support Center for
additional information.
Home
the system.
• Check reagent cartridges for bubbles.
• Perform Reagent Probe to Cartridge and
Reagent Probe Height in Cartridge Alignments.
Refer to step 3 of the
Replace the Old Probe
procedure.