Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x13AD0029
Backup Type on disk
7
doesn't match request
type
0x13AD002A
Instrument serial
7
number on disk doesn't
match Instrument serial
number
0x13AD002B
S/W Version on disk
7
doesn't match MPC S/W
Version
0x13AD0040
Bad Floppy, replace
7
floppy and try Backup
again
0x13AD0041
Chemistry Version
7
number on disk doesn't
match Chemistry
Version number
Event Log
Backup/Restore: Backup type
Incorrect disk format, probably a
on disk doesn't match request
different backup.
type
File: file name
Line: line num
Backup/Restore: Inst serial
Trying to restore from a disk for a
number on disk doesn't match
different system.
Inst serial num
File: file name
Line: line num
Backup/Restore: SW Version
The contents of the backup are from a
on disk doesn't match MPC SW
different version of the Console Software
Version
and cannot be restored on this version.
File: file name
Line: line num
Backup/Restore: Bad floppy,
Cannot check backup.
cannot read from floppy during
backup
File: file name
Line: line num
Backup/Restore: Chemistry
The contents of the backup are from a
Version number on disk doesn't
different version and cannot be restored
match Chemistry Version num
on this version.
File: file
Line: line
Description
• Use another disk if available.
Refer to
• Use another disk if available.
Refer to
• Use another disk if available.
Refer to
• Repeat Backup with different disk.
Refer to
• Use another disk if available.
Refer to
Corrective Action
Backup the
System.
Backup the
System.
Backup the
System.
Backup the
System.
Backup the
System.