UniCel DxC 600/800 Daily Start Up For Training Purposes Only These job aids are shortened versions of procedures found in the reference below. Information in the job aid is correct as of the date published. Verify you have the correct information.
Page 2
Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate the instrument. Beckman Coulter, Inc. urges its customers and employees to comply with all national health and safety standards such as the use of barrier protection. This may include, but is not limited to, protective eyewear, gloves, suitable laboratory attire when operating or maintaining this or any other automated laboratory equipment.
Check Calibration Status Calibrate Process Quality Control using Auto Generation of Control Manually Program/Process Quality Control Sample Processing Low Volume Sample Processing UniCel DxC, Customer, Daily Start Up Job Aid Page 3 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Page 4
UniCel DxC, Customer, Daily Start Up Job Aid Page 4 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Removed, Requested for calibration, or Reserved for Cal. This feature protects you from inadvertently clearing a rack that is needed for one of these conditions. UniCel DxC, Customer, Daily Start Up Job Aid Page 5 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Note: Reagent status can also be viewed on the software Review the Vol (%)/Tests Left, Days Left, and Reagent Status columns for reagent requirements UniCel DxC, Customer, Daily Start Up Job Aid Page 6 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Verify the reagent information is displayed on the monitor Move the cartridge onto the reagent carousel Close the reagent carousel door Additional reagents selected? Procedure complete UniCel DxC, Customer, Daily Start Up Job Aid Page 8 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Verify the reagent information is removed from the monitor. Remove the cartridge Close the reagent carousel door Additional reagents selected? Procedure complete UniCel DxC, Customer, Daily Start Up Job Aid Page 9 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Note: Calibration status can also be viewed on the software Review the Cal Time Left and Cal Status columns to identify reagents requiring calibration UniCel DxC, Customer, Daily Start Up Job Aid Page 10 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Process Quality Control using Auto Generation of Control UniCel DxC, Customer, Daily Start Up Job Aid Page 12 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
QC sample and reset Auto Generation of Control Consult references for Quality Control Procedure complete troubleshooting assistance acceptable? UniCel DxC, Customer, Daily Start Up Job Aid Page 13 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Load the prepared racks on the autoloader with the rack bar code facing to the right Press the <RUN> button on the instrument UniCel DxC, Customer, Daily Start Up Job Aid Page 14 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...
Load the prepared rack(s) on the autoloader with the rack bar code facing to the right Press the <RUN> button on the instrument UniCel DxC, Customer, Daily Start Up Job Aid Page 15 of 15 Rev. F (September 2013) Beckman Coulter Professional Development...