Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x02620036
Level sense error in one
4
or more cartridges
0x0262003A
CC Sample Probe
4
disabled due to
obstruction errors
0x0262003F
Too many tubes not
4
venting correctly,
sampling for the CC
probe will be stopped
Event Log
CAU: Level sense error in one
This error occurs when the reagent
or more cartridges
probes were unable to detect fluid in a
selected cartridge or cartridges during a
Message ID: 0x02620036
reagent load. One attempt by each A
Subsystem: RASched
and B reagent probe is made to level
sense and resulting levels are then
compared for accuracy/consistency.
CAU: CC Sample Probe
Multiple samples had Obstructions
Disabled Due To Obstruction
(Clots).
Errors
Message ID: 0x0262003A
Subsystem: ccScheduler
CAU: CC Sample Probe
Multiple tubes had excessive vacuum in
disabled due to Excess Vacuum
thick stopper closed tubes. CC Sample
Errors
Probe Disabled.
Message ID: 0x0262003F
Subsystem: ccScheduler
Description
• Unload the reagent cartridge and inspect to see
if there are any bubbles in the cartridge
compartments, or residual fluid on the sidewalls
of the openings of the cartridge compartments.
This could cause false triggering of the level
sense. Refer to
Chemistries
• Cartridge must be reloaded to initiate level
sense.
• Check reagent probe alignment for any residual
fluid in wash collar.
Consistent messages may indicate issues with the
level sense function of:
• reagent probe bead assembly
• reagent probe alignments
• level sense board
Contact Beckman Coulter Support Center for
additional information.
• Try flushing probe from the top. Refer to
the Sample and Reagent
• Perform the CC Sample Probe Alignment
(Calibration). Refer to Step 3 of the
Sample and Reagent Probe
• Replace probe. Refer to Step 3 of the
the Sample and Reagent Probe
• Check cap piercer for broken blade. Refer to
Replace CTS Blade/Wick (1-Blade Thick CTS
Option)
CTS Option)
• Perform CC ODC Alignment.
Corrective Action
Load/Unload Cartridge
(CC).
Flush
Probe.
Replace the
procedure.
Replace
procedure.
or
Replace CTS Blade (1-Blade Narrow
to inspect the blade.