Disable Modules; Enable Modules - Beckman Coulter UniCel DxC 600 Instructions For Use Manual

Unicel dxc synchron clinical systems
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

 All programming for Disabled modules is aborted. Modules or subsystems in the Disable mode
will not run, prime, or perform desired maintenance/diagnostic procedures.
When the MC module is enabled, the related reagent is primed 4 times. When the CC subsystem is
enabled, the system runs startup primes.
With an error, for example Reagent too full, the DxC automatically disables a module or subsystem.
The Rgts/Cal status screen changes to Disable (refer to
additional information).
IMPORTANT
the calibration status screen prior to running the system.
IMPORTANT
will not accept work requests at this time. Reagents should be reloaded when a module is Enabled to
ensure that fresh reagents are used for subsequent testing.

Disable Modules

1
Select
Instr Cmd
2
Select
7 Enable/Disable Modules
3
Select the desired module button to disable.
4
Select
OK
5
Select
OK
OR
Select
Cancel

Enable Modules

IMPORTANT
If a reagent is loaded while the module is disabled, prime the module eight times after the module has
been enabled.
Starting from a Stopped state, a module is not enabled until the system is homed and goes to Standby.
1
Select
Instr Cmd
A13914AF
When an MC Module or CC subsystem is enabled, the calibration status will not be lost. Check
When a module is Enabled, the system will go to Standby as it prepares for a prime. The system
from the menu bar.
.
to Disable the desired modules.
to confirm Modules shown in the message box.
to exit without saving changes.
After enabling a chemistry or a module, run controls to verify proper operation.
from the menu bar.
System Status and Commands
Enable/Disable Modules
CHAPTER 4, Reagent Load/Calibration