Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x01FA0016
Primary vacuum level is
4
low.
0x01FA0019
10 psi Air Supply
4
Pressure Is Low.
Event Log
CAU: Primary vacuum is low
This error occurs when the Primary
vacuum drops below 18 in Hg for more
Message ID: 0x01FA0016
than 15 seconds. There is no vacuum
level adjustment now and cannot be set.
CAU: 10 psi Air Pressure Is Low
This error occurs when the transducer
line #129 is outputting a value into the
Message ID: 0x01FA0019
Hydro Board that is below the 10 psi
specification.
Description
• Check vacuum levels and
• Make sure No Foam is present, replace if
necessary. Refer to
Chemistries
• Make sure caps on pressurized bottles are tight.
• Make sure Vacuum Accumulator canister in the
rear of the Hydro is empty.
• Contact Beckman Coulter Support Center for
additional information.
• Check pressure levels and
• Adjust pressure levels if necessary as follows:
— Adjust pressure in Main system above 17
psi.
— Adjust the high pressure system to 17 psi.
— Adjust the low pressure system. The
pressure range for the low pressure system
is 8–11 psi. Set it at 10 psi.
• Contact Beckman Coulter Support Center for
additional information.
Corrective Action
Home
the system.
Load/Unload Modular
(MC).
Home
the system.