Table 12.10 Error Messages and Corrective Actions (Continued)
ID
Pop-up
Class #
0x0262001B
Unrecovered CC
4
Sample Probe
Obstruction
0x0262001D
CC Reagent Probe A
2
Motion Error
Event Log
CAU: Unrecovered CC Sample
This error occurs when there was an
Probe Obstruction
Obstruction present at the CC sample
probe. The probe tried to clear the
Message ID: 0x0262001B
obstruction by flushing, but the
Subsystem: ccScheduler
obstruction could not be removed. The
CC side of the system will finish any tests
that are in progress and then go to a
Stopped state.
CAU: CC Reagent Probe A
This error is caused by a Vertical or
Motion Error
Rotary motion error. This will cause the
instrument to go into a Stopped state.
Message ID: 0x0262001D
Subsystem: RASched
Action Name: action name
Description
• Flush the CC Sample probe per the "As Needed"
Flush the Sample and Reagent Probe
• If flushing probe is not successful,
Sample and Reagent
• If probe is clear and error continues, check for
pinched Sample line tubing.
• If the error continues contact Beckman Coulter
Support Center for additional information.
• Raise canopy and check for obvious obstruction
of the CC reagent Probe A.
• Press the STOP button and
While system is homing, observe motion of the
CC reagent probe A. The probe should complete
its operations in a smooth manner.
• If the error continues, check the event log to
determine the specific subsystem error. Refer to
Event
• If event log indicates a vertical error, make sure
the reagent cartridges on the carousel are
pushed in and seated properly.
• Perform the appropriate vertical or horizontal
CC reagent Probe A alignment. Refer to step 3 in
the
Install the New Probe
• Contact Beckman Coulter Support Center for
additional information.
Corrective Action
procedure.
Replace the
Probe.
Home
the system.
Log.
procedure.