Beckman Coulter UniCel DxC 600 Instructions For Use Manual page 325

Unicel dxc synchron clinical systems
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

4
Loosen the screw at drain top (1). Gently grasp the ISE drain top and remove the top. Do not lose
the attached metal drip screen (2). (The top may require slight wiggling back and forth to
remove.)
Figure 9.68
5
Carefully set aside the drain top (with tubing #22 attached).
6
Remove the drain tube (1) from the drain.
Figure 9.69
7
Using four to five cotton swabs soaked with deionized water, clean the inside of the lower drain
and the top of drain.
A13914AF
Maintenance
As-Needed/As-Required Maintenance