Beckman Coulter UniCel DxC 600 Instructions For Use Manual page 84

Unicel dxc synchron clinical systems
Hide thumbs Also See for UniCel DxC 600:
Table of Contents

Advertisement

System Description
Cartridge Chemistry: Calibration Theory
Table 1.17 Calculation of Calibration Factors for Endpoint and Rate Chemistries (Continued)
Type
Blanked Rate
Chemistries
1-48
FOR HIGH CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate
(Reaction Rate - Blank Rate) = Delta Rate
(Delta Rate
+ Delta Rate
rep1
FOR LOW CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate
(Reaction Rate - Blank Rate) = Delta Rate
(Delta Rate
+ Delta Rate
rep1
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
Sample values are calculated by the following equation:
[(Reaction Rate - Blank Rate) × Cal Factor] + offset = sample value
IMPORTANT
For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
Formula
rep1
rep2
) × 0.5 = Delta Rate
(hi)
rep2
avg
rep1
rep2
) × 0.5 = Delta Rate
(lo)
rep2
avg
(hi)]
avg
A13914AF