Assembly; Visual Inspection; Installing The Breathing System; Inserting The Flow Sensors - Dräger Primus Infinity Empowered Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Assembly

Visual inspection

Inspect all parts for damage and wear, e.g.
cracking, embrittlement or major hardening and
residual soiling.
If necessary, use a dry cloth to remove residual
cleaning agent and disinfectants from the valve
plate, the ventilator module, and the pins of the
breathing system heating.

Installing the breathing system

NOTE
Make sure that all blue rubber seals are correctly
fitted in the bottom section of the breathing sys-
tem.
1 Place the metal valve plate (breathing system
block) (A) on a flat surface.
2 Fit the metal valve plate (B) onto the bottom
section.
Instructions for use Primus Infinity Empowered SW 4.5n
3 Tightly fit the cover.
4 Tighten all five sealing screws (C) a quarter turn
B

Inserting the flow sensors

A
1 Insert the flow sensors (A) with the electric con-
2 Push in the expiratory and inspiratory ports (B),
Cleaning, disinfection and sterilization
D
C
C
clockwise using the key (D) supplied.
B
A
nection in the slot.
with the nose of the port in the slot.
C
C
C
229