JDS Uniphase CellAdvisor JD720C Series User Manual page 66

Cable and antenna analyzers
Table of Contents

Advertisement

Chapter 7 1-Port Cable Loss Testing
3
Figure 35
Amplitude
You can set maximum and minimum values of the Y-scale manually by setting the Top and
Bottom amplitudes or restore the instrument's default full scale defined for each
66
To set the start and stop frequencies, complete the following steps:
a
Select the cell next to the Start Frequency.
b
Enter a value by using the numeric keys or the rotary knob.
c
Select the cell next to the Stop Frequency.
d
Enter a value by using the numeric keys or the rotary knob.
To set the center frequency and the span, complete the following steps:
a
Select the cell next to the Center Frequency.
b
Enter a value by using the numeric keys or the rotary knob.
c
Select the cell next to the Span.
d
Enter a value by using the numeric keys or the rotary knob.
To select a band from the band list, complete the following steps:
a
Touch the icon
The Band List window appears. It displays a list of standard bands stored in the
instrument and custom bands created in the JDViewer. See Figure 35 for example.
b
Highlight your choice from the list, and then tap the Apply button on the Band List
window.
The instrument automatically updates the values in the Start Frequency and Stop
Frequency cells with the selected band's start and stop frequencies.
Tap the Apply button in the lower-right corner of the setup screen to save the change(s).
Band list window
NOTE
Frequency changes after calibration affect the calibration status. If changed within the
registered band, the status turns into CAL ON (I) and it does not require re-calibration. If
changed off the registered band, the status turns into CAL OFF and it requires
re-calibration.
CellAdvisor JD720C User's Manual
next to the band list.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Jd724cJd723clJd726cJd725c

Table of Contents