JD723C, JD724C, and JD726C without 2-por transmission option: Tap the Continue button.
The calibration progress bar appears.
JD725C and JD726C with 2-port transmission option: Tap the O-S-L button.
The calibration progress bar appears.
Disconnect the calibration kit and then connect the SHORT standard of the calibration kit to the
instrument's the RF Out / Reflection port or to the extension cable if used.
Tap the Continue button in the lower-left corner of the screen.
The calibration progress bar appears.
Disconnect the calibration kit and then connect the LOAD (50 Ω) standard of the calibration kit to
the instrument's the RF Out / Reflection port or to the extension cable if used.
Tap the Continue button in the lower-left corner of the screen.
The calibration progress bar appears.
Check the screen to see if the calibration status is now changed to
O-S-L calibration on JD724C using Y-Cal kit
CellAdvisor JD720C User's Manual
CAL
ON.