O-S-L-Thru Calibration Using Electronic Ez-Cal Kit - JDS Uniphase CellAdvisor JD720C Series User Manual

Cable and antenna analyzers
Table of Contents

Advertisement

O-S-L-Thru calibration using electronic EZ-Cal kit

Procedure
1
Connect an extension cable, to the RF Out / Reflection port of the instrument.
2
Connect the EZ-Cal to the end of the extension cable connected to the instrument's the RF Out /
Reflection port.
3
Connect the USB cable to the side of the EZ-Cal and then to the USB Host port to power the
e-calibration kit.
4
Press the CAL hard key.
The on-screen instruction appears to guide you through the e-calibration.
5
Tap the O-S-L-Thru button in the lower-left corner of the screen.
The calibration progress bar appears.
6
Disconnect the EZ-Cal kit and then connect the end of the extension cable connected to the
instrument's RF Out / Reflection port to the instrument's RF In port as illustrated in
Figure 13.
7
Tap the Continue button in the lower-left corner of the screen.
The calibration progress bar appears.
8
Check the screen to see if the calibration status is now changed to
CellAdvisor JD720C Series User's Manual
Chapter 4 Performing Calibration
CAL ON
(E).
41