Chapter 15 Fiber Testing
6
7
8
9
10
Figure 79
122
NOTE
If the image is not properly focused in step 3, you will encounter an analysis failure.
Once the test is completed, check the result displayed on the screen with the following
information:
Image overlay of the zone locations, defects, and scratches
PASS/FAIL result
Specific test result for each Zone
Figure 80 is an example of the test result screen.
Optional. To remove or display image overlay, touch the Overlay screen key.
Optional. If you have checked the box for the Link OPM readings with fiber analysis in the
Setup and you want to integrate OPM readings into a report, complete the following steps:
a
Connect your MP-series optical power meter to your instrument.
b
Tap the Go to OPM screen key.
c
In the optical power meter mode, measure optical powers, and then tap the Store screen
key as needed.
d
Tap the Return to Inspection screen key.
e
Do the following step 9.
Optional. To save the fiber inspection result with specific details as a report in PDF or HTML file
type, complete the following steps:
a
Tap the Save Report screen key.
The Save window appears.
b
Select the file format option from the choices: PDF and HTML.
c
Name the file and save it in the default FiberInspection folder.
See "Using save/load" on page 139 for more information on saving a screen.
d
Tap the Save button.
Touch the Live screen key to return to the live image screen and do more tasks. Otherwise, touch
the Quit screen key to exit.