Quick calibration

The Quick calibration that is available only in the DTF measurement mode is useful when you
want to measure the cable length only, without having to perform a full calibration.
The Quick Cal must be performed after setting the frequencies and connecting an extension
cable. Figure 6 on page 35 illustrates a connection setup with using the port extension cable.
Procedure
1
Connect an extension cable to the RF Out / Reflection port of the instrument.
2
Press the CAL hard key.
The on-screen instruction appears. It contains the Quick Cal button.
3
Connect the OPEN standard of the calibration kit to the instrument's the RF Out / Reflection port
or to the extension cable if used.
4
Tap the Quick Cal button in the lower-left corner of the screen.
The calibration progress bar appears.
5
Check the screen to see if the calibration status is now changed to
Figure 12
Quick calibration on JD724C
NOTE
CAL ON (Q) is valid only for a DTF measurement in the single layout. If you change the
display mode to others, calibration status goes back to the previous one.
CellAdvisor JD720C Series User's Manual
Chapter 4 Performing Calibration
CAL ON
(Q).